B. Operating Environment
Please make sure that the environment is adequate for the InBody770 installation. This equipment is designed
forindooruse.Ifinstallingoutdoors,thefollowingrequirementsmustbefullled.
C. Installation Instructions
1. Open the packing box of the InBody770 and remove the packing pads. Then take the InBody770 out of the box.
Temperature range 50~104˚F(10~40˚C)
Relative humidity 30 ~ 75% RH
Atmospheric pressure 70 ~ 106kPa
Polyethylene foam cover
Packing pads
*
Please refer to the following illustrations to properly transport the equipment.
•
If you have any problems installing your InBody770, please contact InBody for assistance.
•
Do not transport the equipment by holding the screen portion or the joints of the hand electrodes.
•
Keep the packing materials provided for repacking the equipment in the future. Other wastes should be
disposed of according to relevant laws and regulations.
Caution
•
Si vous avez des problèmes lors de l'installation de votre InBody770, veuillez contacter InBody pour
obtenir de l'aide.
•
Ne transportez pas l'équipement en tenant l'écran ou les articulations des électrodes à main.
•
Conservez les matériaux d'emballage fournis pour remballer l'équipement si nécessaire. Éliminer les
autres déchets conformément aux lois et règlements pertinents.
Attention
•
Using the InBody770 on carpet may cause static electricity, which could damage the equipment. If installing
the InBody770 on carpet is unavoidable, please use an antistatic mat.
•
Install the InBody770 on a leveled, non-vibrating surface. Installing the equipment on an uneven surface
may cause the examinee to fall down. Test results may also be inaccurate.
•
Never clean the hand and foot electrodes with liquid spray or detergent directly. The equipment may corrode
and/or malfunction if the liquid or detergent leaks inside. Use the InBody Tissue when cleaning the InBody770.
*
For inquires regarding the InBody Tissue, please contact InBody.