EasyManua.ls Logo

Incon Auto-Learn TS-LS300 - Page 6

Incon Auto-Learn TS-LS300
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PREFACE Page P-3
P
LS300 User’s Guide
Each product line must be precision tested and verified free of leaks before any component of the
LS300 system is installed. After verification, the LS300 Auto-Learn
®
system is installed, wired, and
calibrated. The supplied leak-generating device/calibration kit* is attached to a needle valve, and the
appropriate line channel selector switch is placed in the learn position at the console. Each line is
calibrated separately, one at a time, to “learn” the parameters and characteristics of a line with a
known induced leak-rate.
At each STP housing, the needle valve kits* are assembled and installed in the pressure test ports,
and the LSU300 pressure-transducers are installed in 2 inch NPT line leak detector ports (LSU300
cable length must not exceed 500 feet).
After a line is calibrated the leak-generator device is moved to the next product line for Auto-Learn
calibration. The entire process is repeated until all lines are separately calibrated and the console is
left in the Run – Detect mode.
The front panel has a power on light and 4 status-lights per line. The status lights show currently
running line leak tests, faults, and alarms. Fault-alarms will cause indicator lights to blink or flash.
Certain alarms will also activate the internal alarm horn. Manual line leak tests may be started and
Figure P-1 LS300 LLD System Components
2 CHNL LS300
* The needle valve kit and the leak generating device/calibration kit have not been
evaluated by Underwriters Laboratories
.
LEAK GENERATING
DEVICE
NEEDLE
VALVE
LSU300 TRANSDUCER
# 3 with
3/4 Chnl
# 4 with
4 Chnl
PRODUCT
ID
STICK-ON
DANGER
3 OR 4 CHNL
NO-STRIP
SPLICE
2 CHNL

Related product manuals