9
Name
Hosting
Number
Data
Type
Length
Read/
Write
Description
Upper limit of measure-
ment
0X 00 0A
Floating
Point
2
Read/
Write
Upper limit of measured value
(20mA corresponding value)
Lower limit of
measurement
0X 00 DC
Floating
Point
2
Read/
Write
Lower limit of measurement value
(4mA corresponding value)
Upper temperature limit 0X 00 OE
Floating
Point
2
Read/
Write
Upper temperature limit
Lower temperature limit 0x 00 10
Floating
Point
2
Read/
Write
Lower temperature limit
Measured value offset 0x 00 12
Floating
Point
2
Read/
Write
Adjust measurement
Temperature offset 0x 00 14
Floating
Point
2
Read/
Write
Adjust temperature value
Damping
coefficient
0x 00 16 Integer 1
Read/
Write
0-10
Device address 0x 00 19 Integer 1
Read/
Write
1-255
Baud rate 0X 00 1A Integer
1
Read/
Write
0=2400 , 1=4800 , 2=9600
3=19200, 4=38400
Restore factory 0X 00 1B Integer 1 Write
ORP calibration value 0x 00 30
Floating
Point
2
Read/
Write
Calibrating slope 0x 00 34 Integer 2 Read -0.1984
Zero point
calibration solution
0x 00 36 Integer 1
Read/
Write
0=7.00, 1=6.86
Slope calibration solu-
tion
0x 00 38 Integer 1
Read/
Write
0=1.68 , 1=4.01 , 2=9.18
3=10.1 , 4=12.45
Manual
temperature
0X 00 3A
Floating
Point
2
Read/
Write
25°c
Zero confirmation 0X 00 3E Integer 1 Write
Slope confirmation 0X 00 3F Integer 1 Write
Measured ADC 0x 00 66 Integer 1 Read