MISURATORE DI PRESSIONE
DIGITALE DA BRACCIO
INN-014
MANUALE D’USO
USER MANUAL
INN-014_manuale_260117 ese.indd 1 26/01/17 10:49

Questions and Answers

  • D
    Donna AndersonAug 6, 2025
    What to do if INNOLIVING INN-014 Medical Equipment shows error code E6?
    • J
      John EllisonAug 7, 2025
      If the INNOLIVING Medical Equipment displays error code 6 (E6), this indicates a low battery. Replace the batteries to resolve this issue.
  • K
    Kenneth HarrisonAug 11, 2025
    What to do if INNOLIVING INN-014 Medical Equipment shows error code E1?
    • Y
      ymurphyAug 11, 2025
      If the INNOLIVING Medical Equipment displays error code 1 (E1), it means that the cuff is not fastened properly or is not in the correct position. Ensure the cuff is applied correctly and sit comfortably and still, keeping the cuff at the same level as your heart.
  • B
    brownbonnieAug 15, 2025
    How to fix a blank display on INNOLIVING INN-014 Medical Equipment?
    • A
      Ann HigginsAug 15, 2025
      If nothing appears on the display of your INNOLIVING Medical Equipment, even when it's turned on, the batteries may be drained. Replace all batteries with new ones. Also, check that the battery polarities are correctly aligned, with the negative and positive ends matching the indications in the battery compartment; if they aren't, re-install them correctly.
  • M
    mzhangAug 20, 2025
    What to do if INNOLIVING Medical Equipment shows error code E2?
    • C
      Craig DavisAug 20, 2025
      If the INNOLIVING Medical Equipment displays error code 2 (E2), it means you moved your arm or body during the measurement. To resolve this, make sure you remain very still and quiet during the measurement.
  • L
    langkyleAug 23, 2025
    What to do if INNOLIVING Medical Equipment shows error code E4?
    • D
      Daniel WardAug 25, 2025
      If the INNOLIVING Medical Equipment displays error code 4 (E4), it means the unit does not measure or there is a measuring error. If you have a very weak or irregular heartbeat, the device may have difficulty determining your blood pressure. Also, sit comfortably and still, and fasten the cuff again carefully.
  • C
    Crystal WatersAug 28, 2025
    What to do if INNOLIVING INN-014 Medical Equipment shows error code E3?
    • J
      Jason BernardAug 28, 2025
      If the INNOLIVING Medical Equipment displays error code 3 (E3), the cuff may not be applied correctly. Check whether the tube connection of the cuff is secured to the unit properly.
  • R
    Rebecca WelchSep 2, 2025
    What does error code E5 mean on INNOLIVING Medical Equipment?
    • L
      Laura YuSep 2, 2025
      If the INNOLIVING Medical Equipment displays error code 5 (E5), the measurement range is over 300mmHG. It is recommended to see doctor as soon as possible.

INNOLIVING INN-014 Specifications

General IconGeneral
CategoryMedical Equipment
ModelINN-014
BrandINNOLIVING
TypePulse Oximeter
SpO2 Measuring Range70% - 99%
DisplayOLED
Power2 x AAA batteries
PR Measuring Range30 BPM - 240 BPM