Manual
DefibrillationwithPaddles
Switch on the device
Connect paddle connector at the front
of the defibrillator
Select Manual Mode
Check that there is no sync mode at the
top of the screen
Select energy level with the +/- buttons
on the Sternum paddle
Use gel on the paddles, place the
paddles on the patient and hold them
stable
Press charge button on Apex paddle
and wait until the charging is
completed. Do not move or remove
the paddles
Push the two Shock buttons on both
paddles simultaneously
Asynchronous
Manual
CardioversionwithPaddles
Switch on the device
Connect paddle connector at the front of
the defibrillator
Select Manual Mode
Check the expiry date of the ECG
electrodes and attach them to the patient
Connect ECG cable at the front of the
defibrillator and connect the lead wires to
the ECG electrodes
Push Sync On soft button
Check that sync mode is displayed at the
top of the screen
Set up Latched or Non-Latched mode with
the Menu buttons in the Sync Menu
Select the appropriate ECG lead with good
quality signal. Observe the 'S' sync
markers on the screen above the ECG
curve
Select energy level with the +/- buttons on
the Sternum paddle
Use gel on the paddles, place the paddles
on the patient and hold them stable
Press charge button on Apex paddle and
wait until the charging is completed. Do
not move or remove the paddles!
Push and hold down the two Shock buttons
on both paddles simultaneously, until the
shock is delivered. Do not move or
remove the paddles, nor release Shock
buttons until the shock is delivered!
Synchronized
Cardio-Aid 360-BQUICKGUIDE
1
3
4
6
9
2
10
8
5
7
S
ternum
S
t
e
r
n
u
m
ApexA
p
e
x
Usingwith electrodesPaddle
6
11
InnomedMedicalInc.
SzabóJ.u.12,H-1146
Budapest,Hungary
R-6789-1/A
0120
11
3
4
5
6
7
11
3
9
4
10
11
5
6
7
Question and Answer IconNeed help?

Do you have a question about the Innomed Medical Cardio-Aid 360-B and is the answer not in the manual?

Innomed Medical Cardio-Aid 360-B Specifications

General IconGeneral
BrandInnomed Medical
ModelCardio-Aid 360-B
CategoryMedical Equipment
LanguageEnglish

Summary

Manual Asynchronous Defibrillation with Paddles

Initial Device and Mode Setup

Switch on device, connect paddle connector, select Manual Mode, and verify no sync mode is active.

Energy Selection and Paddle Application

Select energy level, apply gel to paddles, and place them on the patient's chest.

Charging and Shock Delivery

Press charge button until complete, then simultaneously push both shock buttons to deliver the shock.

Manual Synchronized Cardioversion with Paddles

Initial Device and ECG Setup

Switch on, connect paddles/ECG, select Manual Mode, attach electrodes, and enable sync mode.

Sync Mode Configuration and Signal Optimization

Configure sync mode (Latched/Non-Latched) and select ECG lead with good signal quality.

Energy Selection and Paddle Application

Select energy level, apply gel to paddles, and place them on the patient's chest.

Charging and Shock Delivery

Press charge button until complete, then hold both shock buttons until the shock is delivered.

Manual Asynchronous Defibrillation with Patches

Initial Device and Mode Setup

Switch on device, connect patch cable, select Manual Mode, and attach patches to patient.

Energy Selection and Patch Application

Select energy level and press the action button to initiate charging.

Shock Delivery

Press the action button on the front panel to deliver the shock to the patient.

Manual Synchronized Cardioversion with Patches

Initial Device and ECG Setup

Switch on, connect patch cable, select Manual Mode, attach patches, and enable sync mode.

Sync Mode Configuration and Signal Optimization

Configure sync mode (Latched/Non-Latched) and select ECG lead with good signal quality.

Energy Selection and Patch Application

Select energy level and press the action button to initiate charging.

Charging and Shock Delivery

Press action button for charging and hold until shock is delivered.