Innovo® CMS 50F PLUS User Manual
29
Guidance and manufacture’s declaration – electromagnetic
immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic
immunity
The CMS50F PLUS Pulse Oximeter is intended for use in the
electromagnetic environment specified below. The user should
assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Complianc
e level
Electromagneti
c environment
- guidance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
6 kV
contact
8 kV air
6 kV
contact
6 kV air
Floors should
be wood,
concrete or
ceramic tile. If
floor are
covered with
synthetic
material, the
relative
humidity should
be at least 30%.
The
manufacturer
may
recommend the
ESD
precautionary
procedures to
user.