Innovo® CMS 50F PLUS User Manual
Recommended separation distances between portable and
mobile
RF communications equipment and the EQUIPMENT or
SYSTEM –
for EQUIPMENT or SYSTEM that are not
LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the
CMS50F PLUS Pulse Oximeter.
The CMS50F PLUS Pulse Oximeter is intended for use in an
electromagnetic environment in which radiated RF disturbances
are controlled. The user can help prevent electromagnetic
interference by maintaining a minimum distance between
portable and mobile RF communications equipment
(transmitters) and the CMS50F PLUS Pulse Oximeter as
recommended below, according to the maximum output power
of the communications equipment.
Rated
maximum
output
power of
transmitter
(W)
Separation distance according to frequency of
transmitter
(m)