EasyManua.ls Logo

Integra FIREBOY Series - Connecting a tubing; Electricity supply and connecting a switch; Operating; Working with liquids

Integra FIREBOY Series
4 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
144952_V01_QSG_FIREBOY_EN.docx 3/4
Connecting a tubing
A propane/butane gas cylinder must be equipped with a pressure regulator to not exceed
the maximum input pressure of 50 mbar.
Connect the tubing from the central gas supply or from the gas
cylinder to the previously installed adapter and fix it with a clip, if
necessary.
Electricity supply and connecting a switch
Insert the mains plug in the socket on the rear side of the FIREBOY
and plug it to the electricity mains. (FIREBOY plus can also be
operated with the optional battery.) Switch on the device.
Insert the plug of the foot or benchtop switch in the socket on the
rear side, if required.
Operating
There a several options for flame ignition:
FIREBOY plus
Ignition
Burning time
SENSOR
Object passes the sensor
window (two times or once)
Timer with the option to extinguish by
pressing the MENU KNOB
BUTTON
By pressing the MENU KNOB
(continuous mode)
Extinguish by pressing the MENU KNOB,
maximum burning time defined
FOOTSWITCH
By pressing the foot or
benchtop switch
As long as pressed or fix by timer with the
option to extinguish by pressing the switch
or MENU KNOB
FIREBOY eco
Ignition
Burning time
Button
By pressing the MENU KNOB
(continuous mode)
Extinguish by pressing the MENU KNOB,
maximum burning time of 60 min
Foot/benchtop
By pressing the foot or
benchtop switch
As long as pressed
Working with liquids
In order to protect the burner chamber from contamination when
working with liquids, the folding stand at the base can be used to
incline the FIREBOY.
The inclination direction can be modified by removing the folding
stand after pressing it together and inserting it on the opposite side.
Maintenance
Allow the burner head to cool off before cleaning the device. Switch off the device and
disconnected it form the gas supply. Allow the device to dry before operating it again.
Clean the FIREBOY with a cloth moistened with soapy water or with a 70 % ethanol solution.
If the burner head get soiled, it can be pulled vertically out for disinfection.
If the FIREBOY has been contaminated, wipe it with a lint-free cloth lightly soaked with a
suitable disinfectant and wipe dry directly.
Check the tubing frequently for brittle or porous areas. Replace it, if damaged, or at the latest 8
years after production (year imprinted on the tubing).