8
by the system or affect the performance of the system and miss low/high blood
glucose incidents.
Continuous glucose monitoring cannot be used as a basis for diabetes diagnosis,
such as for insulin injection treatment. It cannot replace blood glucose meter.
The value of blood glucose may differ from the reading of glucose. Using
glycemic readings as a diagnostic basis for diabetes may result in low/high
blood glucose.
Do not ignore the symptoms of high / low blood glucose levels. If the sensor
glucose reading does not match the symptoms, a blood glucose meter should be
used to measure the blood glucose even if the CGM reading is within the
normal range.
Calibrate optionally, or when the sensor glucose readings might be inaccurate
and you might miss a low or high blood glucose value.
In rare cases, the sensor electrode may fracture. If the sensor electrode breaks
and there is no visible sensor electrode on the skin, do not attempt to remove it
yourself. Seek professional medical help in the condition of infection or
inflammation - redness, swelling or pain. If the sensor electrode ruptures, please
report to our technical support.
Do not use sensors when its sterile packaging is damaged. Using unsterilized
sensors may lead to infection.
The storage temperature of the sensor is 2
o
C-25
o
C. The sensor can be
transported at room temperature. The storage time is the shelf life (8 months) of
the sensor. If the temperature range of the refrigerator is 2
o
C-25
o
C, the sensor
can be stored in the refrigerator. Improper storage results in inaccurate blood
glucose readings and missing low/high blood glucose levels. If the sensor is
stored in the refrigerator, please take it out half an hour before use. The sensor
can be used at room temperature.
The device is not designed to remain in the body long-term, or to be replaced
‘immediately’ with a similar or identical device.
Your transmitter communicates with your App via Bluetooth. The
communication might be affected by strong electro-magnetic field, keep your
CGMS away from any strong electro-magnetic field. Otherwise, degradation of
the performance of this equipment could result in malfunctioning of the system.
With the prompt function turned off, the App cannot give a prompt
(notification) even if the Transmitter and the App are within communication
range.
Symptoms that may be caused by high/low blood glucose should not be
ignored. If the symptoms you are experiencing do not match the CGMS
readings, or you suspect that the CGMS readings may be inaccurate, check the
glucose by performing a fingerstick blood test using a blood glucose meter.
When CGMS shows that your glucose level is low or near the low limit,
perform a fingerstick blood test using a blood glucose meter.
The CGMS contains small parts that may cause choking once swallowed, keep