Recommended minimum separation distances
RF wireless equipment is used in various healthcare locations therefore when the
CGMS is used in close proximity to other medical equipment and/or systems, the
medical equipment and/or systems’ basic safety and essential performance may be
affected. Continuous Glucose Monitoring Systems has been tested with the immunity
test level in the below table and meet the related requirements of IEC 60601-1-2:2014.
The customer and/or user should help keep a minimum distance between RF wireless
communications equipment and Continuous Glucose Monitoring Systems as
recommended below.
EMC Statement
Test
frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
power
(W)
Distance
(m)
Immunity
test level
(V/m)
385
380-
390
TETRA
400
Pulse
modulation
18Hz
1.8
0.3
27
450
430-
470
GMRS 460
FRS 460
FM
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704-
787
LTE Band
13, 17
Pulse
modulation
217Hz
0.2
0.3
9
745
780
810
800-
960
GSM
800/900,
TETRA
800,
iDEN 820,
CDMA
850,
LTE Band
5
Pulse
modulation
18Hz
2
0.3
28
870
930
1720
1700-
1990
GSM
1800;
CDMA
1900;
GSM
1900;
DECT;
LTE Band
1, 3,
4, 25;
UMTS
Pulse
modulation
217Hz
2
0.3
28
1845
1970
2450
2400-
2570
Bluetooth,
WLAN,
802.11
b/g/n,
RFID
Pulse
modulation
217Hz
2
0.3
28