Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the iRestore Professional,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3
V
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Recommended Separation Distance
d=1.2P
d=1.2P 80 MHz to 800 MHz
d=2.3P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitter,as determined by an
electromagnetic site survey
a
should be
less than the compliance
level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
a
Field strengths from fixed transmitter, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters
, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the iRestore Professional is used exceeds the applicable RF compliance level above, the iRestore Professional should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the iRestore Professional.
b
Over the frequency range 150 kHz to 80 MHz,field strengths should be less than 3V/m.
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The iRestore Professional is intended for use in the electromagnetic environment specified below. The customer or the user of the
iRestore Professional should assure that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not
apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.