Troubleshooting
No power light
↓
Is the line cord completely plugged
into the wall?
→ No→
Plug line cord into the wall.
↓
Yes
↓
Is the wall outlet working properly?
→ No→
Try another wall outlet.
↓
Yes
↓
Is the fuse blown?
→ Yes→
Replace fuse.
↓
No
↓
Contact Iris Sample Processing or
an Authorized Repair Facility.
Centrifuge will not spin, or shuts
off prematurely
↓
Is the cover closed and latched?
→ No →
Completely close the cover.
↓
Yes
↓
Is the error light flashing?
→ Yes →
Identify source of error from flashing
indicator lights, correct the error condition,
and push the Stop/Open button to clear the
error.
↓
No
↓
Contact Iris Sample Processing or
an Authorized Repair Facility.
Centrifuge will not open at end
of cycle
↓
Is the bowl gasket completely
seated and sealing the bowl?
→ NO→
Press the bowl gasket all the way down into
the housing slot.
↓
YES
↓
Contact Iris Sample Processing or
an Authorized Repair Facility.