9. Product Specifications
9.1 Delivery form
Programat S1
Power cord
Temperature Checking Set S1 Starter Kit
Operating Instructions
Sinter tray
Sinter tray fork
USB download cable
Equipment Service Passport
Various accessories
9.2 Technical data
Power supply 118–240 V / 50–60 Hz
Overvoltage category II
Contamination level 2
Acceptable voltage fluctuations +/- 10%
Max. power consumption 16 A at 118 V
8 A at 240 V
Electrical fuses 250 V / T16 A high breaking (heating circuit)
Dimensions electrical fuses Diameter 5 x 20 mm
Dimensions of the closed furnace
Depth: 430 mm / Width: 310 mm / 390 mm (with cooling tray)
Height: 570 mm
Usable size of the sintering chamber Diameter 80 mm
Height 80 mm
Max. firing temperature 1530 °C/2786 °F
Weight Furnace base: 10 kg
Furnace head: 17 kg
Safety information
The sinter furnace complies with the following guidelines:
IEC 61010-1:2001
EN 61010-1:2001
UL 61010-1:2004
CAN/CSA-C22.2 No. 61010-1:2004
IEC 61010-2-010:2003
EN 61010-2-010:2003
CAN/CSA-C22.2 No. 61010-2-010:2004
Radio protections / electromagnetic compatibility EMV tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5°C to +40°C (+41°F to +104°F)
Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31°C
(87.8° F) gradually decreasing to 50 % relative humidity
at 40°C (104°F); condensation excluded.
Acceptable ambient pressure:
The furnace is tested for use at altitudes of
up to 2000 m above sea level.
9.4 Acceptable transportation and storage conditions
Acceptable temperature range -20 to +65°C (-4°F to +149°F)
Acceptable humidity range Max. 80% relative humidity
Acceptable ambient pressure 500 mbar to 1060 mbar
Use only original packaging of the Programat S1 together with the
respective foam material for shipping purposes.