JIF-2002 WEIGHING INDICATOR&CONTROLLER
38
C.Err 4The calibration mass has been mistakenly entered as a
value greater than the maximum capacity.
Please reduce the weight of calibration mass, and re-enter the
weight value.
C.Err 5: The calibration mass has been wrongly entered zero or it is
smaller than the minimum capacity.
Please increase the weight of calibration mass, and re-enter the
weight value.
C.Err 6: The load cell output is too low.
Replace your load cell with a more sensitive one or adjust the
minimum division.
C.Err 7: The load cell signal pins are reversed, or the load cell output
voltage is too low.
Check the load cell connections if reversed or load cell failure.
C.Err 8: The load cell output voltage at maximum capacity is too high.
Check the load cell specification or load cell failure.
C.Err 9: The maximum, capacity has been wrongly entered as a value
smaller than 100.
Check Resolution Table.
C.Err10 :The maximum, capacity has been wrongly entered as a value
greater than 750,000.
Check the load cell specification or load cell failure.