23
7.5.3 Calibration Setting
Note that the items for setting differ depending on the reference materials used for calibration.
Item Description
Air saturated water
This calibration uses air-bubbled water as the reference. Enter
[Atmospheric Pressure] at the time of calibration.
RMDO (Reference Material DO)
(Under development)
A reference material can be used for calibration, which is
under development as of May 2016.
displayed on the label of the reference material into the
[RMDO] and [Salinity].