T r oubleshooting
IM-459000
12
TROUBLESHOOTING
Symptom
Power indicator does not
light up.
Power indicator lights up
but actuator does not
run. Relays in control box
are heard clicking.
Power indicator lights up
but actuator does not
run. No relay noise is
heard from control box.
Control box completely
dead on battery and no
relay clicking is heard.
Control box okay apart
from one direction on one
channel.
Possible Cause
Not connected to mains
The fuse has blown
Control box defective
Actuator plug not pushed
into control box properly
Actuator defective
Control box defective
Control box defective
Handset defective
Handset plug not inserted
properly
Emergency stop engaged
Charger plugged in
Battery completely flat
Battery defective
Handset defective
Control box defective
Action
Connect to mains
Replace fuse, if the sys-
tem is prepared for exter-
nal fuse replacement, or
send the system for repair
Send Control box for
repair
Push actuator plug into
control box properly
Replace actuator
Replace control box
Send control box for
repair
Send handset for repair
Check all cord connections
Deactivate emergency stop
Unplug charger
Charge battery
Replace battery
Send handset for repair
Send control box for
repair

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Joerns Healthcare Hoyer P-C-HLA-2 and is the answer not in the manual?

Joerns Healthcare Hoyer P-C-HLA-2 Specifications

General IconGeneral
BrandJoerns Healthcare
ModelHoyer P-C-HLA-2
CategoryLifting Systems
LanguageEnglish

Related product manuals