To Previous Page IconTo Previous Page
Loading...
9
ANCHOR INTRASPINAL CATHETER TO THE FASCIA
Attach the silicone anchor to the catheter where the
catheter enters the lumbo-dorsal fascia.
Fasten the anchor to the fascia using heavy non-
absorbable suture.
Reconfirm catheter patency with saline flush or CSF
backflow. Replace the SNAPLOCK
cap to prevent
CSF leakage.
Verify and document catheter position using fluoroscopy.
ALTERNATIVE TECHNIQUE (DUAL ANCHOR):
Attach the cylindrical silicone anchor to the catheter
where the catheter enters the lumbo-dorsal fascia.
Fasten the anchor to the fascia using heavy non-
absorbable suture.
Create a strain relief loop in the catheter of
approximately 3 cm.
Attach the right angle silicone anchor to the catheter and
fasten to the fascia using heavy non-absorbable suture.
Reconfirm catheter patency with saline flush or CSF
backflow. Replace the SNAPLOCK
cap to prevent
CSF leakage.
Verify and document catheter position using fluoroscopy.
MEDSTREAM
PROGRAMMABLE INFUSION SYSTEM Surgical Technique Guide
492192.QX7:Layout 1 4/14/09 4:09 PM Page 9