CONNECT THE MEDSTREAM PUMP CATHETER
TO THE SURESTREAM CONNECTOR
Slide the small diameter end of the strain relief sleeve
approximately 4 cm over the pump catheter.
Slide the end of the pump catheter onto the pin
connector with a slight twisting motion.
Slide the strain relief sleeve over the pin connector until it
contacts the stop.
Secure the strain relief sleeve onto the catheter and the
connector by tying 2 ligatures of heavy non-absorbable
suture.
Test for leaks and catheter patency by administering a
bolus injection of preservative free saline through the
bolus port.
Note:
Refer to the MEDSTREAM OR Poster for
detailed bolus instructions and illustrations.
14