What to do if Johnson & Johnson Laboratory Equipment displays H2O2 sensor fault?
V
Veronica CampbellSep 5, 2025
If the Johnson & Johnson Laboratory Equipment displays 'H2O2 SENSOR FAULT', it indicates that the hydrogen peroxide sensor is out of the expected range. Clean the lamp and detector optical windows. Run diagnostics. Verify operation of the UV lamp. Replace any failed components.
P
pattersonstevenAug 25, 2025
What to do if Johnson & Johnson Laboratory Equipment says collection box is full?
J
Jason DiazAug 25, 2025
If your Johnson & Johnson Laboratory Equipment displays 'COLLECTION BOX FULL, PLEASE CHANGE BOX', it indicates that either the collection box has two empty cassettes or the collection box is not present. Resolve this by emptying the collection box or inserting it properly. If the issue continues, you may need to troubleshoot or replace the collection box switch.
P
Patricia BatesAug 25, 2025
What to do if Johnson & Johnson Laboratory Equipment displays temperature has not risen?
E
Erica TurnerAug 26, 2025
If the Johnson & Johnson Laboratory Equipment displays 'TEMPERATURE HAS NOT RISEN', it means the operating temperatures were not reached during warm-up, possibly because the door is open, or a heater or sensor has failed. First, ensure the sterilizer door is closed and wait 30 minutes. If the problem persists, run diagnostics and replace any failed components.
K
Kevin HobbsSep 13, 2025
What to do if Johnson & Johnson Laboratory Equipment displays 5 volt supply out of range?
M
Mr. Jeremy GarciaSep 14, 2025
If the Johnson & Johnson Laboratory Equipment displays '5 VOLT SUPPLY OUT OF RANGE', the 5 VDC supply is out of tolerance. Run the diagnostics. If the diagnostics fail, replace the ATXDC power supply.
G
garnermichaelSep 23, 2025
What to do if Johnson & Johnson Laboratory Equipment displays vacuum detected, run diagnostics?
G
greenestephenSep 23, 2025
If the Johnson & Johnson Laboratory Equipment displays 'VACUUM DETECTED, RUN DIAGNOSTICS', a sensor has detected a fault during operation. Run diagnostics. Replace failed component.
J
Justin GoldenSep 9, 2025
What to do if Johnson & Johnson Laboratory Equipment displays H2O2 bulb detector fault?
R
Rachel WilliamsSep 10, 2025
If the Johnson & Johnson Laboratory Equipment displays 'H2O2 BULB/DETECTOR FAULT', the hydrogen peroxide sensor is out of the expected range. Inspect operation of UV lamp; replace lamp if failed. Clean the optical windows of the lamp and detector. Run diagnostics. Replace failed component.
P
pateldavidSep 1, 2025
What to do if Johnson & Johnson Laboratory Equipment displays condenser temperature too high?
M
Monica RasmussenSep 1, 2025
If the Johnson & Johnson Laboratory Equipment displays 'CONDENSER TEMPERATURE TOO HIGH', it means the condenser temperature was higher than expected, possibly due to a sensor or heater failure. Run diagnostics and adjust or replace the vaporizer/condenser assembly.
D
David HollandSep 23, 2025
What to do if Johnson & Johnson Laboratory Equipment displays vent timeout?
R
rhondacontrerasSep 23, 2025
If the Johnson & Johnson Laboratory Equipment displays 'VENT TIMEOUT', the time to vent the chamber is excessive. Run diagnostics. Troubleshoot or replace failed component.
M
mariamartinSep 18, 2025
What to do if Johnson & Johnson Laboratory Equipment displays IPC failure cancellation?
W
williamsjulieSep 18, 2025
If the Johnson & Johnson Laboratory Equipment displays 'IPC FAILURE CANCELLATION', a generic system error has occurred. Power sterilizer down and turn on again. If problem persists, run diagnostics. Replace failed component.
C
Chad TerryAug 28, 2025
What to do if Johnson & Johnson Laboratory Equipment displays chamber 1 temperature too low?
M
Michelle WoodsAug 28, 2025
If the Johnson & Johnson Laboratory Equipment displays 'CHAMBER 1 TEMPERATURE TOO LOW', it indicates the chamber front temperature was lower than expected, possibly due to a sensor or heater malfunction. Run diagnostics and replace the failed component.