JPD-FR403
Infrared Thermometer
Instruction Manual
Shenzhen Jumper Medical Equipment Co., Ltd
Manual Version: 1.0
Date of Issue: 2021/05
0482

Questions and Answers

  • B
    Barbara BrewerAug 7, 2025
    Why Jumper JPD-FR403 Thermometer fails to power on?
    • J
      Jacob HayesAug 7, 2025
      Your Jumper Thermometer might not be powering on due to several reasons: the batteries may be low (in this case, change the batteries), the polarities of the batteries might be reversed (make sure that the batteries are installed correctly), or the thermometer itself could be damaged (in this case, contact the manufacturer).
  • J
    Jessica JonesAug 17, 2025
    What to do if temperature reading is lower than normal on Jumper JPD-FR403 Thermometer?
    • R
      Rebekah HudsonAug 17, 2025
      If your Jumper Thermometer shows a temperature reading lower than the typical body temperature range, it could be due to a few reasons. The lens of the temperature probe might be dirty, so try cleaning it with a cotton swab. Alternatively, the thermometer probe might not be aligned correctly with the eardrum; reposition it to ensure proper alignment. Also, if the thermometer was recently moved from a cold environment, wait more than 30 minutes before using it to take a measurement.
  • M
    Matthew FrazierAug 26, 2025
    What does "Er1" mean on my Jumper JPD-FR403?
    • B
      Brian GreenAug 27, 2025
      If your Jumper Thermometer displays "Er1", it means the ambient temperature is either lower than 10°C (50.0°F) or higher than 40°C (104°F). To resolve this, take a measurement when the ambient temperature is between 10°C (50.0°F) and 40°C (104°F).

Summary

Unpacking Check

Package Contents

Safety Precautions

Warning

FCC Compliance Statements

US Radiation Exposure Statement

Symbols

Body Temperature Basics

Product Description

Overview

Infrared Thermometer JPD-FR403 measures body temperature based on infrared energy emitted from the eardrum or forehead.

Structure

The thermometer consists of a shell, an LCD, a measure button, a beeper, an infrared temperature sensor, and a Microprocessor.

Operating principle

The infrared temperature sensor collects infrared energy emitted by the eardrum or skin surface.

Intended use

Measures human body temperatures; forehead mode for all ages, ear mode for >3 months old.

Contraindications

Do not use if the ear is infected with otitis or suppuration.

Features

Safety and Accuracy

Highlights good safety features and high measurement accuracy.

User Experience

Focuses on easy operation and quick measurement for user convenience.

Versatile Functions and Modes

Covers diverse functions, broad application scope, and child mode operation.

Product Structure

Display Description

Sounds and Backlight Color Instructions

Display and Operating Instructions

Out of Measuring Range Display

Displays 'HI' or 'Lo' for readings outside the normal temperature range.

Switching Between Mute and Un-mute

Instructions for toggling the sound on or off and the mute symbol display.

Switching Between Object and Body Temperature

Procedure for switching between object and forehead/ear measurement modes.

Switching Between °F;°C

Instructions on how to change the temperature units from Fahrenheit to Celsius and vice versa.

Measuring Forehead Temperature

Measuring Object Temperature

After a Measurement

Guidance on recalling measurements and cleaning the thermometer after use.

Recall 20 memories

Error information & low battery

Error Code Erl

Indicates ambient temperature is too high or too low.

Error Code ErC

Indicates data read/write error or incomplete temperature correction.

Low Battery Alert

Alerts when battery voltage drops below a critical level.

Measuring Ear Temperature

Replacing Batteries

Cleaning and Disinfection

Cleaning

Recommended detergents and steps for cleaning the thermometer probe and body.

Disinfection

Recommended disinfectants and steps for disinfecting the thermometer.

Maintenance

Preventive Inspection and Maintenance

Covers regular safety checks, cleaning, storage, and battery removal for long-term storage.

Troubleshooting

Specifications

Security Class

Storage and Transportation

EMC Information-Guidance and Manufactures Declaration

Electromagnetic Emission Guidance

Provides declarations and guidance regarding electromagnetic emissions for the device.

Electromagnetic Immunity Guidance

Provides declarations and guidance regarding electromagnetic immunity for the device.

Electromagnetic Immunity for Non-Life-Supporting Systems

Details immunity requirements for systems not supporting life, including ESD and magnetic fields.

Recommended RF Separation Distances

Specifies minimum distances for RF communications equipment to ensure EMC compliance.

Warranty and After-Sale Service

Jumper JPD-FR403 Specifications

General IconGeneral
Response Time1 second
DisplayLCD
Power Supply2 x AAA batteries
Automatic Shut-offYes
TypeInfrared Thermometer
Measurement Range32°C - 42.9°C (89.6°F - 109.2°F)
Accuracy±0.2°C (±0.4°F)