Hangzhou Junce Instruments Co., Ltd.
current remaining power percentage before using it for the first time. If the
effective capacity of the battery is unknown or the current remaining capacity is
not known, it needs to be tested according to the following steps:
1 Empty the battery pack or put it until you don’t want to discharge it.
2 Enter the preset battery capacity setting in the system settings, and
set the capacity value as large as possible (for example, set the estimated
20Ah to 30Ah);
3 Set the remaining capacity percentage to 0%, and then charge the
battery pack;
4 Check the remaining capacity value displayed after fully charged, and
set this value to the preset battery capacity;
i. Fine adjustment of ambient temperature
Click on the ambient temperature, the "Input:Temperature calibration"
pop-up box will pop out, enter the corresponding value for fine-tuning, the
default temperature calibration is 100. The ambient temperature is shown in
Figure 3-1-19 below.
j. Ambient temperature unit switching
Click the ambient temperature value to pop up the pop-up "Temperature
Unit", select Celsius or Fahrenheit, the default is Celsius. Figure 3-1-20
Temperature unit as shown below.
Figure 3-1-10 Output
status is off
Figure 3-1-11 Over
voltage protection status
Figure 3-1-12 Voltage
calibration