Details the product's high sensitivity, stable signal, high precision, wide measurement range, and good waterproof performance.
Lists key technical specifications including power supply, output signal, response time, temperature range, and humidity range.
Provides a checklist of devices to verify before proceeding with the installation of the sensor.
Explains the wide voltage input power interface and crucial notes on RS485 signal cable polarity and addressing.
Details the wiring sequence and terminal connections for the economical version of the sensor.
Illustrates the wiring connections for the sensor's LCD display shell, showing terminal assignments.
Describes the wiring sequence for sensors with an external probe, specifying line colors for power and communication.
Covers the physical installation methods for different sensor models, including rail mount and wall mount options.
Guides users on how to connect the sensor transmitter to a PC using an RS485 to USB adapter.
Explains the usage of the configuration software for testing and configuring sensor parameters, including connecting and setting options.
Details fundamental communication settings such as Modbus RTU protocol, data bits, parity, stop bits, and baud rate.
Lists Modbus register addresses for humidity, temperature, device address, and baud rate, along with their read/write permissions.
Provides a practical example of Modbus communication, including inquiry frames for reading temperature and humidity data.