Sensor removal
Switch off the DTR to cut off power from the sensor.
Close wash valve.
1. Unscrew and remove sensor nameplate.
2. Loosen the cable gland.
3. Unscrew the wires.
4. Remove the sensor cable and cable gland.
Note: If another in-line sensor is connected to the
same DTR, disconnect the loose cable from the DTR