Do you have a question about the KCI infoV.A.C and is the answer not in the manual?
Crucial safety warnings and information accompanying the V.A.C.® Therapy System.
Legal disclaimer regarding KCI product warranties and limitations of liability.
Essential conditions and precautions for proper and safe use of KCI products.
Identifies and illustrates the various components and features of the InfoV.A.C.® Therapy Unit.
Provides an overview of the manual's purpose and scope for servicing the unit.
Details initial steps for unpacking, inspecting, and preparing the unit for patient use.
Recommended protocols, standards, and general guidelines for infection control and cleaning.
Step-by-step instructions for cleaning the Therapy Unit and its power supply.
Procedure to visually inspect the unit for any physical damage or wear.
Checks the functionality and condition of the hanger arm and anti-slip pads.
Details battery check, change, and charging procedures for the unit.
Verifies power-on, screen readability, and procedures for data transfer.
Procedures for inspecting canister bellows and ensuring proper canister fit.
Checks and sets the correct time and date displayed on the unit.
Verifies pressure transducer accuracy and performs sensor calibration tests.
Steps for unit preparation and accessing engineering software screens for diagnostics.
Verifies the unit pump responds correctly to pressure setting adjustments.
Ensures the unit correctly detects, displays, and sounds alarm conditions.
Confirms test completion, sets defaults, and ensures the battery is fully charged.
Guidelines for accompanying documentation and safe transport of the unit.
Instructions for repairing or replacing the rubber block on the hanger arm assembly.
Procedure for regluing or repairing the anti-slip pads on the unit's rear housing.
Lists the physical, electrical, and environmental specifications of the InfoV.A.C.® Therapy Unit.
Provides a list of available spare parts with their corresponding part numbers.
Explains the meaning of various symbols and icons used in the manual and on the unit.
Provides contact details for KCI for support, supplies, and information.
Details safety testing related to ground continuity and earth leakage for compliance.
| Therapy Type | Negative Pressure Wound Therapy (NPWT) |
|---|---|
| Portability | Portable |
| Pressure Settings | Continuous, Intermittent |
| Power Source | AC power, Internal rechargeable battery |
| Alarms | Canister Full, Low Battery |
| Canister Capacity | 300 mL, 500 mL, 1000 mL |
| Display | LCD Screen |