Technical data
General information
Protection type: IP20
Measurement range: 30° to 150°
Minimum measurement distance: 20 mm to the surface
Resolution: 0.01°
Measurement accuracy: ±0.2°
*)
Repeat accuracy: 0.1°
Power supply: 5 V DC, 500 mA
Battery: 2 integrated LiPo batteries with safety circuits
Nominal battery voltage: 3.7 V
Battery capacity: 2 x 330 mAh
*)
See the operating instructions for a measurement accuracy reference.
Ambient conditions
Operating temperature: +5 °C to +40 °C
Charging temperature: +5 °C to +40 °C
Storage temperature: -20°C to +45°C
Relative air humidity: 5% to 95% (non condensing)
Laser
Laser class: 2 in accordance with IEC 60825-1:2014
Wavelength: 660 nm
Laser output: 7 mW (pulsed)
Laser output calibration uncertainty: ±0.9 mW
Laser beam diameter: 3 mm (at the exit point)
Beam divergence: 35 mrad
Pulse frequency: 167 Hz
IEC 60825-1:2014, Table 10, Condition 3:
200 mm (most restrictive location); 36 µs (impulse
duration in orifice plate)
For a complete list of the technical data, please refer to the operating instructions or
the data sheet.