Fill-O-Meter Instruction Manual
Page 5 of 6 www.KegLand.com.au Last Updated 5/10/2020 3:59 PM
3. Calibration
1. Press MENU 3 times and the factory setting (009.40 LPM) will be displayed as the
instantaneous value
2. Press SET for 3 seconds and use the up and down arrows to set the pin number to 1234
(default password)
3. Use the up or down arrows to change the LPM calibration
4. Press MENU to save the changes
The factory default K-factor is 9.4 LPM.
To calculate the Frequency, use the following formula:
Frequency = K-factor (L/min) x time (min)
= 9.4 x 60
= 564 pulses/L/min