OPERATION
73
Cycle Step Start: Determines the starting step for the Cycle
Repeat field. Range: 1~999. The Cycle Step Start field must be less
than the Cycle Step End field.
Cycle Step End: Determines the ending step for the Cycle Repeat
field. Range: 2~1000. The Cycle Step End field must be greater
than the Cycle Step Start field.
Line 2:
Step Header: This field contains the “Stepheader text. Required.
Do not remove from this location.
Time(sec) Header: This field contains the “Time(sec)” header text.
Required. Do not remove or alter from this location.
Voltage (V) Header: This field contains the “Voltage (V)” header
text. Required. Do not remove or alter from this location.
Current (A) Header: This field contains the “Current (A)” header
text. Required. Do not remove or alter from this location.
Line 3+:
Step Number: The Step Number field numbers each step in the
test function. It has a range of 1~1000. All steps must be
numbered in numerical order. No step number can be “skipped”
between the first and last step.
Time(sec): Designates the duration time of the corresponding
step in seconds. It has a range of 0.05 seconds to 20 days. It has a
step resolution of 0.05 seconds (50ms). A setting of 0 will skip the
step.
Voltage (V): Designates the voltage of the corresponding steps in
volts. MIN or MAX can be used to set the voltage to minimum or
maximum, respectively.
Current (A): Designates the current of the corresponding steps in
amps. MIN or MAX can be used to set the current to minimum or
maximum, respectively.