SECTION3
IEEE-488
Reference
Programming
Example
3-18
+
ing
the source value.
6.
Use
the sweep delay to compensate for any settling time required
by
the external
circuit connected to the source-measure unit.
7.
If
a pulse sweep
is
modified, it's delay times
will
no
longer
be
recalculated for
changes in autorange measure, filtering, integration time,
and
default delay.
8.
Depending
on
your
controller
and
program, a
bus
timeout
may
occur during a
long sweep
delay.lfyou
trigger a reading
and
ask for it right away, the
bus
is
held
off until the measurement
is
available, possibly causing a timeout.
9.
Error
and
warning conditions for the A command are described
in
the
U1
error
status
word
and
U9
warning status
word
(see paragraph 3.6.19).
10
20
30
40
50
60
REMOTE716
OUTPUT
716;''F0,1X"
OUTPUT716;"Q1,2,10,2,2,100X"
OUTPUT
716;" A-2,2,2X"
OUTPUT
716;"A,150,4,5X"
END
+
! Set for IEEE-488 communication
!
Source
V,
measure
I,
sweep
! Linear stair sweep, five
2V
steps,
11Vor15Vrange, 100msecdelay
!
Change point 2 to
-2V
!
Change points 4
and
5 to
150msec
delay

Other manuals for Keithley 237

Related product manuals