SECTION3
IEEE-488
Reference
Default
Description
3-48
+
pulses -
Quantity of pulses in a pulse sweep waveform
(0-500)
range-
Selects the source range in
all
sweep types:
1-source V-source
0=
Auto Auto
1
=
InA
2=
lanA
3=
IDOnA
4=
lJJA
I.
IV
(236,
237); 1.5V
(238)
11
V
(236,
237);
15V
(238)
nov
1100V(237)
5 =
IOJJA
6=
IOOJJA
7=
lmA
8=
lOrnA
9=
lOOmA
10=
1A(238)
*
*
*
*
*
*
*illegal
V -source option.
start, stop
-Starting
and
stopping source values for stair waveforms.
!-source
(236,
237)
(238)
V-source (236,238)
(237)
-100.00
to
+lOO.OOmA
-1.0000
to
+
l.OOOOA
-110.00
to
+llO.OOV
-1100.0
to
+llOO.OV
step-
Incremental absolute value for stair sweeps. Maximum value
is
twice the full
scale range
of
the
highest range
used
in generating the sweep.
!-source (236,237)
(238)
V-source (236,238)
(237)
-200.00
to
+200.00mA
-2.0000
to
+2.0000A
-220.00
to
+220.00V
-2200.0
to
+2200.0V
toN-
Duration of sweep level
in
milliseconds
(0-65000)
toFF-
Duration of bias level in milliseconds
(0-65000)
Upon
power-up,
or
after receiving a DCL
or
SOC
command, the instrument defaults
to the last programmed selections. All items
in
the sweep buffer are preserved.
A
JOX
command restores factory default values of
IV
/lmA
for level/start,
IOV
/lOmAforstop,lV
/lmAforstep,llV
/lOmAforrange,Omsecfordelay,lOfor
count/pulses,
10 log points
per
decade,
and
lOOmsec
for
toN/toFF.
All items in the
sweep buffer are cleared.
The
Q
command creates
or
appends a sweep list
by
defining the parameters of a se-
lected sweep type. The sweep types include: fixed level, linear stair, logarithmic
stair, fixed level pulsed, linear stair pulsed,
and
logarithmic stair pulsed.
+

Other manuals for Keithley 237