8.1 ADJ Models
8.1.1 Automatic adjustment using a built-in weight
Automatic adjustment is performed:
having disconnected the scales from mains,
having detected a change in temperature by 2°C from the previous adjustment,
after 3 hours from the previous adjustment.
The adjustment is initiated only when after turning the scales it shows zero
indication for 30 minutes (no changes in load, no pressing of buttons, and
no changes in external conditions).
Ensure menu setting <C1-0>, see chapter 11.2.
Procedure:
In case of the adjustment error (e.g. object on the plate) the following error
message will be displayed: <CAL NO>, repeat the adjustment procedure.