ADB_ADJ-BA-e-2032 25
11.2 Menu overview
Menu item Settings Description
Adjustment mode
C1 - 0
Automatic internal adjustment (ADJ models)
C1 – 1
External adjustment
(ADB models: always use this setting)
Note: At the models with internal adjustment
weight the adjustment with external weight is not
recommended..
C1 - 2
Manual internal adjustment (ADJ models)
Reference quantity
C2 – 0*
10
C2 - 1 20
C2 - 2 50
C2 - 3 100
C2 - 4 1000
Automatic
zero point correction
C3 - 0 No Zerot Tracking
C3 – 1*
1D Zero-point Tracking
C3 - 2
2D Zero-point Tracking
C3 - 3 3D Zero-point Tracking
C3 - 4 4D Zero-point Tracking
C3 - 5 5D Zero-point Tracking
C3 - 6 Not documented
Baud rate
C4 - 0
1200
C4 - 1*
2400
C4 - 2
4800
C4 - 3 9600
Data output C5 - 0 Automatic output of stable weighing values
C5 - 1 Via remote control command W (w)
C5 - 2 Continuous data output
C5 – 3*
Output for stable and instable weighing values
after pressing PRINT key
Sound by pressing
the button
C6 – 0 switched on
C6 - 1*
switched off
Automatic internal
adjustment interval
(only ADJ models)
C7 - 0*
2 hours
C7 - 1
3 hours
C7 - 2 4 hours
C7 - 3 6 hours
C7 - 4 8 hours
Filter C8 - 0* Low interference degree
C8 - 1
Medium interference degree
C8 - 2
High interference degree
C8 - 3 Not documented
* = factory setting