5. Within three seconds you will be booted to FreeDOS and have a command prompt.
6. Change to the folder that contains the BMC bits. In this example type “cd bmc1v8a” and press
Enter.
The next step starts the 20 second window of vulnerability. If you remove power or
pull the USB flash drive during this 20 second window, your M9037A will most
likely lock
-up and will not boot. To get your M9037A working again you must send
it to Keysight for repair.
7. Type “upbmc” and press Enter to start the update progress.
8. The update takes approximately 20 seconds. When the update is complete, the M9037A
automatically powers-down. You are done with Step 2 and ready for Step 3.