Getting Started
STEP 5. Verify Operation of the Network Analyzer Modules
Figure 1-11 Operator’s Check Dialog Box
If the Operator’s Check Fails
1. Clean the test ports, and the open(s) or short(s). Torque to specification.
Repeat the check.
2. If the check still fails, return the network analyzer module to Keysight. See
“How to Return an Instrument for Service” on page 7.