Keysight X-Series Signal Generators Error Messages 41
Error Messages
List of Error Messages
This error should not appear with normal Arb operation. This is a
hard fault if it should occur. If this error is seen for external
source data from the LVDS interface, then the interface and its
associated timing and control is suspected.
-240 Hardware error; BBG Arb SDRAM initialization failure: {name}.
There is a problem initializing the BBG waveform memory (SDRAM).
Report this error to service office.
-240 Hardware error; BBG Arb SDRAM refresh failure.
The baseband board BBG waveform memory failed a refresh cycle.
Report this error to service office.
-240 Hardware error; BBG Arb waveform memory (SDRAM) cannot be
accessed.
There is a problem accessing the BBG waveform memory (SDRAM).
Report this error to service office.
-240 Hardware error; BBG Communication failure
Communication with the BBG FPGA failed; specifically, the SSIO bus
failed to respond. Cycle the instrument's power. Report this
failure to service office.
-240 Hardware error; BBG DMA error: {name}.
An internal error occurred during DMA transactions to the Baseband
assembly. Please report this error to a service office.