1.3.3 FUNCTION AND LEAK TESTS
Following overhaul, the King Nutronics Corporaon PN 3666-12-101 Control Valve Assembly must be tested for
proper funcon and leakage in accordance with the following procedure:
1. Congure the Model 3666 Automac Pressure Calibrator for standard gauge calibraon.
2. Install the dust plug into the Control Unit Quick Disconnect TEST port.
3. Program the following test points using psi units in the upscale and downscale direcons: 20; 200; 2,000;
5,000; and 10,000. Operate the unit through the programmed range.
4. Program a new test with a 20-psi test point. Press the [ADV] key to commence tesng.
5. Wait for the MANUAL SET message to be displayed. Aer this message appears, observe the ACTUAL pressure
in the display for stability.
a. If the displayed value connues to increase past the programmed setpoint, the “pressure side” of the
valve is leaking.
b. If the displayed value decreases, apply a bubble soluon to the VAC port to check for leakage. If leakage
is evident, the “vent side” of the valve is leaking.
1-4