EasyManua.ls Logo

Lab-Line 314 - Page 22

Default Icon
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
ALARMS, 314/316 SERIES AUTOMATIC CO2 INCUBATORS:
POWER OUTAGE: In the event of a power outage, the LED display will flash to
alert the user to take corrective action. If the power outage is of a temporary
nature, pressing any key will restore unit to operating status. Everything operates
normally after a power outage.
DOOR AJAR WARNING: In case the door is inadvertently left open beyond a 5
minute period, an audible beep is sounded and is synchronized with the flashing
of the DOOR-HEAT indicator light to warn the user to correct the condition.
AUDIBLE AND VISUAL ALARMS: The following table summarizes the audible
and visual alarms that are initiated upon the occurrence of various events.
AUDIBLE FLASHING FLASHING
BEEP LIGHT LED
ABOVE KEY DISPLAY
Temp (exceeds limits): YES YES NO
CO2 (exceeds limits): YES YES NO
Door Ajar: YES YES NO
Power Out: NO NO YES
Press any key to silence alarm that will remain silent for a period of one
hour. However, the flashing light will continue. If the condition initiating the alarm
remains uncorrected at the end of that time, the audible alarm will resume.
PASSWORD, CONTROL PANEL, 314/316 SERIES AUTOMATIC CO2 INCUBATORS:
In the event that you want to prevent unauthorized access to the
operation of the unit, a password feature is available
ADDITIONAL INCUBATOR FEATURES, 314/316 SERIES AUTOMATIC CO
2
INCUBATORS:
CO2 sensor (on thermal-conductivity units only).
Mechanically convected airflow.
1.2 cubic feet, 304-stainless steel chamber; full-view heated glass door;
2 shelves