29
SECTION 6
OPERATION
DANGER: DO NOT USE IN THE PRESENCE OF FLAMMABLE OR
COMBUSTIBLE MATERIALS OR EXPLOSIVE GASES. DO NOT USE IN THE
PRESENCE OF PRESSURIZED OR SEALED CONTAINERS. FIRE OR
EXPLOSION MAY RESULT, CAUSING DEATH OR SEVERE INJURY.
WARNING: DO NOT HEAT ANY SUBSTANCE ABOVE A TEMPERATURE
THAT WILL CAUSE IT TO EMIT TOXIC FUMES—DEATH OR SEVERE
INJURY MAY RESULT.
DANGER: USE ONLY CARBON DIOXIDE IN THE INCUBATOR. DO NOT
UNDER ANY CIRCUMSTANCE INJECT OXYGEN OR OTHER EXPLOSIVE
GAS OR MIXTURE INTO UNIT. FAILURE TO OBSERVE THESE
PRECAUTIONS CAN RESULT IN EXPLOSION AND/OR FIRE AND SERIOUS
INJURY OR DEATH TO PERSONNEL AND PROPERTY DAMAGE
START-UP, 314/316 SERIES AUTOMATIC CO
2 INCUBATORS:
Connect unit to power supply meeting the requirements as noted on the
nameplate.
Turn the power switch ON.
OPERATING BASICS, 314/316 SERIES AUTOMATIC CO2 INCUBATORS:
Press the SET key first and then the parameter—CO2, TEMP, DOOR
HEAT—which is to be adjusted.
Use the UP and DOWN arrow keys to enter the desired value.
Wait 5 seconds and the value is entered automatically into the unit's
nonvolatile memory. Storage is confirmed via two short beeps at the end
of 5 seconds.
If you access a parameter and initiate no action within the time frame of 10
seconds, the unit times out and returns to the prior state.
Each press of the SCAN/HOLD key is programmed to follow a sequential
pattern as follows: HOLD Temperature (for continuous viewing), HOLD
CO
2 (for continuous viewing) and alternate scanning of temperature and
CO2.
To view your previously established set point for any parameter, press
SET and the TEMP, CO2 or DOOR HEAT key(s).