TEMPERATURE CALIBRATION, 314/316 SERIES AUTOMATIC CO2 NCUBATORS:
The Incubator is calibrated at the factory. However, if required, the
temperature readout may be calibrated by pressing the following keys
in sequence: CAL, TEMP.
Use of a calibrated indicating device positioned in the approximate
geometric center of the chamber is critical to this procedure.
Use the Up and Down arrow keys to correct the displayed temperature on
the LED readout to match your chamber temperature reading. Wait 5
seconds and the adjustment is self-entering