TEMPERATURE CALIBRATION, 314/316 SERIES AUTOMATIC CO2 NCUBATORS:
The Incubator is calibrated at the factory. However, if required, the
temperature readout may be calibrated by pressing the following keys
in sequence: CAL, TEMP.
• Use of a calibrated indicating device positioned in the approximate
geometric center of the chamber is critical to this procedure.
• Use the Up and Down arrow keys to correct the displayed temperature on
the LED readout to match your chamber temperature reading. Wait 5
seconds and the adjustment is self-entering