a hazardous condition and may damage the centrifuge. Individual tubes, strip tubes
and/or adapters must always be loaded symmetrically to ensure proper balance. Never
run the centrifuge with only one strip tube in place.
Safety Precautions:
NEVER use the centrifuge in any manner not specified in these instructions.
NEVER
operate the centrifuge without a rotor properly attached to the shaft.
NEVER
fill tubes while they are in the rotor. Liquid spillage may harm unit.
NEVER put hands in the rotor area unless the rotor is completely stopped.
NEVER move the centrifuge while the rotor is spinning.
NEVER use solvents or flammables near this or other electrical equipment.
NEVER centrifuge flammable, explosive or corrosive materials
NEVER centrifuge hazardous materials outside of a hood or proper containment facility
ALWAYS load the rotor symmetrically. Each tube should be counterbalanced by another
tube. Weight of individual tubes may not exceed of 3.0g.
ALWAYS locate the centrifuge within easy access to an electrical outlet.
ALWAYS
use only microcentrifuge tubes made from plastic and designed to withstand
centrifugal forces of at least 2,000 xg.
Cleaning:
Before using any cleaning or decontamination methods except those recommended by
the manufacturer, check with the manufacturer that the proposed method will not
damage the equipment. To clean the centrifuge, use a damp cloth and a mild,
noncorrosive detergent (ph <8). After cleaning, ensure all parts are dried thoroughly
before attempting to operate the unit. Do not immerse the centrifuge in liquid or pour
liquids over it.
Questions and Service:
Should you have a question about the centrifuge or require service for the unit, please
use the contact numbers listed below. Do not send in a unit for service without first
calling to obtain a repair authorization number and a decontamination form. Should the
centrifuge require service, be sure to decontaminate it and enclose the decontamination
form. The unit should be properly packed to avoid damage. Any damage resulting from
improper packaging shall be the responsibility of the user.
Specifications: Spare Parts:
Dimensions... ........ 6” x 6” (153 x 153mm) Replacement 1.5ml rotor: C-1300-RT
Electrical ............... 115V, 50/60Hz,130mA Replacement strip rotor: C-1300-RTS
Fuse ..................... 0.5A, 250V (F) Replacement allen key: C-1300-RW
Speed ................... 6000rpm Replacement fuses: C-1200-F
RCF ...................... 2000 x g Ambient Operating Range: 0 to 40°C,< 85%
Relative Humidity
For indoor usage only
Over voltage category II
Pollution degree 2