DigiBlock User Manual
www.labtechsrl.com
Page 16 of 25
iTouch Separate CONTROLLER
LabTech iTouch controller operates by simply touching the keys on the screen.
Connect the iTouch controller to the back of DigiBlock main unit, switch ON the DigiBlock and the iTouch screen is
automatically ready for operations.
1. Temperature control range: room temperature to 210°/415°C (based on different DigiBlock model).
2.
Working time: from 1 to 15 heating steps. Each step can be set from -2 to 9999min.
-2” Maximum heating speed to reach temperature setpoint.
-1” End the program.
0” Maintain previous step time setting.
1) Temperature sensor mode: built-in K type sensor or optional PT100 external sensor.