Editing an Application Setup Interface
Edit a soil digestion method in Soil category, heat from room temperature to 100℃ and hold in 100until
switch off the unit.
A. In home page, touch Setup to enter method set interface.
B. Enter Method Edit interface, touch Soil and Method1
C. Step 01: Touch the Temp. Text box, input 100 via the digital key. Then touch Time text box to input
-2 which means heat in max. heating speed (Normally, the average heating speed is around
5℃/min).
Step 02: Touch the Temp. text box and input 100 via the digital key. Then touch Time text box to
input 0 which means hold the temperature at previous step setpoint.