- 8 -
4.2. Power and safety device
Over Temp
Over temperature protection device
- It is a dial type temperature controller. It automatically cuts the heater output off when there is
over temperature due to equipment malfunction.
- Set this 10~20% higher than the set temperature.
- Over temperature protection device is analog type; therefore, it will not accurately match the
displaying temperature.
Power switch
Power on/off switch
Circuit Breaker
Device to supply power and to cut off an overcurrent