Fill sample bottle (0689) with water sample and replace cap.
Fill clean tube (0205) to the 5 mL line with water sample and cap.
Insert the sample tube into the ColorQ 2x meter.
Turn meter on, press 'Blank' to calibrate, then proceed to FCL.
Add 5 drops of DPD 1A and 5 drops of DPD 1B to the same tube.
Cap the tube and invert 3 times to mix the solution.
Insert tube and press button to read 'FCL' for Free Chlorine.
Press button to advance to 'TCL' for Total Chlorine.
Remove cap from reacted FCL tube and add 5 drops of DPD 3.
Cap with pink cap, invert 3 times to mix, and insert tube.
Press button to read 'TCL' for Total Chlorine.
Press button past 'Br' to advance to 'pH' reading.
Fill clean tube (0205) to 5mL with water sample and add 5 drops of pH.
Cap with red cap, invert 3 times to mix, and insert tube.
Press button to read the pH value of the sample.
Press button to select 'Alk' for Total Alkalinity.
Fill tube to 5mL, add 1 ALK 2X IG tablet, crush, cap green.
Invert tube 10 times to mix and insert into ColorQ 2x.
Press button to read 'Alk' for Total Alkalinity.
Press button to select 'Ca Hard' for Calcium Hardness.
Fill tube to 5mL, add 1 CA HARD 2X IG tablet, crush, cap purple.
Invert tube 3 times to mix and insert into ColorQ 2x.
Press button to read 'Ca Hard' for Calcium Hardness.
Press button to go to 'CYA' for Cyanuric Acid.
Fill tube to line, add 1 CYANURIC ACID IG tablet, crush, cap grey.
Invert 3 times to mix, let tube sit for TWO minutes.
Insert tube and press button to read 'CYA' for Cyanuric Acid.
Press button to return to blank meter or hold until displayed.