EasyManua.ls Logo

LaMotte TRACER pH PockeTester 1741 User Manual

LaMotte TRACER pH PockeTester 1741
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
TRACER
POCKETESTER
TM
pH
CODE 1741
Question and Answer IconNeed help?

Do you have a question about the LaMotte TRACER pH PockeTester 1741 and is the answer not in the manual?

LaMotte TRACER pH PockeTester 1741 Specifications

General IconGeneral
pH Range0.0 to 14.0 pH
pH Resolution0.1 pH
pH Accuracy±0.2 pH
Automatic Temperature CompensationYes
DisplayLCD
WaterproofYes, IP67
Temperature Resolution0.1°C
Temperature Accuracy±0.5°C
Power Supply2 x 1.5V AAA batteries
Battery Life>500 hours

Summary

Introduction and Specifications

Introduction

Congratulations on purchasing the TRACER pH PockeTester, a revolutionary measurement device offering ease of use and accuracy.

Specifications

Details the technical specifications of the TRACER pH PockeTester, including display, operating conditions, and accuracy.

Parts and Accessories

Meter Description and Operation

Front Panel Description

Identifies and describes the components located on the front panel of the TRACER meter.

TRACER Display Overview

Explains the various readings, indicators, and bar graph elements shown on the TRACER's LCD screen.

Basic Operation

Powering and Electrode Recognition

Instructions for turning the TRACER on/off and how it recognizes connected electrodes.

Automatic Calibration

Details the process for automatic electrode calibration to ensure accurate pH readings.

Temperature Unit and Data Hold

Guide to changing temperature units (°C/°F) and using the Data Hold function.

Display Messages

Explains status indicators like 'BAT', 'HOLD', and 'CAL Reminder'.

pH Testing Procedures

RENEW Indicator

Explains the 'RENEW' warning indicating the probe may be nearing the end of its life.

pH Testing Overview

Defines pH, the pH scale, and how pH relates to acidity and alkalinity.

pH Display Interpretation

How to read pH values and interpret the bar graph movement on the display.

Getting Started and Buffer Prep

Initial steps for electrode setup and preparing pH buffer solutions.

pH Calibration and Measurement

pH Calibration

Step-by-step guide for calibrating the TRACER meter using 1, 2, or 3 point methods.

pH Measurement

Instructions for taking pH measurements in liquid samples.

pH Measurement of Soil

Procedure for measuring pH directly from soil samples.

Data Management and Maintenance

Storing and Recalling Readings

How to store up to 15 readings and recall them later.

Storage Recommendations

Proper methods for storing the electrode and meter to preserve performance.

Battery Replacement

Detailed steps for replacing the CR2032 batteries in the meter.

Electrode Cleaning and Conditioning

Guidance on cleaning the electrode surface with ethanol or isopropyl alcohol.

Troubleshooting and Electrode Replacement

Replacing the Electrode

Instructions for safely removing and attaching electrodes to the TRACER meter.

Troubleshooting Common Issues

A table listing common problems, their causes, and recommended solutions.

Expandability and Warranty

Expand Your TRACER

Information on interchangeable electrodes for Total Chlorine and ORP measurements.

Warranty Information

Details the manufacturer's warranty for the instrument and probe.

Related product manuals