Issue 1 10/2019 Page 2 of 6
EL-SIE-2+
High Accuracy Temperature and Humidity
USB Data Logger, EasyLog Cloud Compable
NO SOFTWARE TO INSTALL
All the soware needed to congure your EL-SIE-2+, and view and analyse the data it logs, is contained within the product itself. Just
connect the logger to your PC or Mac with a USB cable, open any web browser and in the address bar type “hp://EasyLog.local”. No
internet connecon is needed, and you can save this address in your bookmarks or favourites as normal.
The interface is easy to use and allows the following parameters to be congured:
Logger and channel names
Measurement units
Logging rate and start mode
Up to 16 separate alarms (high/low/pre-alarm/cumulave)
with thresholds, delay and hold
Display and LED indicator modes
Once the logger is running, you can plug it back
into your computer and see the latest data, device
status and the event log. You can also chose to
stop the logger and change the conguraon, or
just let it connue logging.
EASYLOG CLOUD DATA STORAGE
Store your data securely, and make it available from
any internet-connected PC or mobile device, with
EasyLog Cloud. The EL-SIE-2+ can upload logged data
to the Cloud from your PC or Mac, making sharing
and analysis easier than ever. For more informaon
and to set up an account on EasyLog Cloud, visit
www.easylogcloud.com.
Display Logger Status Explanaon
USB Connected The logger is connected via the USB cable
Push to Start The logger is set up for Push to Start logging, a long
press of either buon will start logging
Delayed Start The logger is set up for Delayed Start logging and will
automacally start logging at the specied me
Triggered Start The logger is set up for Triggered Start logging and
will start logging as soon as the specied limit (either
temperature, humidity or pressure) is reached
Low Baery The baeries are low and should be changed when
possible
Memory Full The memory is full and logging has stopped
Memory 90% The memory is 90% full, and data should be down-
loaded when possible
Calibraon Pend-
ing
The calibraon will expire in < 30 days
Calibraon
Expired
The calibraon has expired
Logger Running The logger is sll logging but can be stopped by a long
press of the boom buon
Alarm Triggered An alarm is currently acve on the logger
Cumulave Alarm A cumulave alarm is acve on the channel currently
being displayed
High Alarm A high alarm is acve on the channel currently being
displayed
Low Alarm A low alarm is acve on the channel currently being
displayed
Held High Alarm A high alarm is being held on the channel currently
being displayed
Held Low Alarm A low alarm is being held on the channel currently be-
ing displayed
High Pre-alarm A high pre-alarm is acve on the channel currently
being displayed
Low Pre-alarm A low pre-alarm is acve on the channel currently be-
ing displayed