What to do if ATT error appears on Leuze electronic AMS 348i Analytical Instruments?
- TTracy WilsonAug 17, 2025
If the ATT (insufficient received signal level) error appears on your Leuze electronic Analytical Instruments, it could be due to several reasons. The reflector might be soiled, so try cleaning the reflector or the glass lens. Another reason could be performance reduction due to environmental conditions such as snow, rain, fog, condensing vapor, or heavily polluted air. In this case, optimize usage conditions. Also, check the alignment to ensure the laser spot is fully on the reflector and remove any protective foil from the reflector.