EasyManuals Logo

LEYBOLD TRIVAC B User Manual

Default Icon
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
10
-1
h
.
m
-13
2
10
-5
Saugvermögen/Pumping speed/débit
S65B
S40B
D40B
D65B
468
10
-4
10
-3
10
-2
10
-1
10
0
10
1
10
2
10
3
mbar
10
1
10
2
2
4
6
8
10
0
Druck/Pressure/pression
Description
7
Fig. 5 Pumping speed characteristics of the TRIVAC S 40 B to TRIVAC D 65 B, 50 Hz operation, SI units
m
3
· h
-1
m
3
· h
-1
mbar
mbar
mbar
DN
l
IP
S 40 B
46
40
< 2.5 · 10
-2
< 6.5 · 10
-1
65
40 KF
2.3/3.7
54
D 40 B
46
40
< 1 · 10
-4
< 5 · 10
-3
40
40 KF
1.7/2.6
54
S 65 B
75
65
< 2.5 · 10
-2
< 6.5 · 10
-1
65
40 KF
2.5/3.3
54
50 Hz operation, SI units
D 65 B
75
65
< 1 · 10
-4
< 5 · 10
-3
40
40 KF
2.0/3.3
54
Nominal pumping speed*
Pumping speed*
Ultimate partial pressure
without gas ballast*
Ultimate total pressure
with gas ballast*
Water vapour tolerance*
Connection ports
Oil filling min./max.
Type of protection
* as per DIN 28 400 and following numbers
1.6 Technical data
with gas ballast
without gas ballast

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LEYBOLD TRIVAC B and is the answer not in the manual?

LEYBOLD TRIVAC B Specifications

General IconGeneral
BrandLEYBOLD
ModelTRIVAC B
CategoryWater Pump
LanguageEnglish