Lighthouse Worldwide Solutions ApexZ Operators Manual
248083466-1 R6 32
Identify your monitoring point location
and select that location on the Sample
Map View screen by pressing that
location name button. The selected
location button will highlight with a
yellow boarder and the Start button will
highlight green indicating the ApexZ
instrument is ready to sample at the
current location. Make sure your sample
tube and sample probe are in place at
the selected location point.
Press the green start button to start
sampling at the current selected
location.
After sampling completes then move the ApexZ to the next monitoring point. Make sure your sample tube and
probe are in place at the new location and then select the matching current location on the Sample Map View
screen. Press the green start button to start sampling at the new location.
Location Color Status
Yellow boarder = current location
Green back ground = sample completed
with no warning or alarms exceeded.
Blue back ground = currently sampling.
Note if a sample has multiple cycles then
the back ground color will change after
the first cycle is completed if the particle
count exceeds the warning or alarm
thresholds or the sensor goes into fault.
Grey back ground = location not sampled
Orange back ground = sample completed
at that location and the warning threshold particle count was exceeded.
Red back ground = sample completed and the alarm threshold particle count was exceeded at that location.
Note a red back ground color will also display if the instrument was in fault for its laser, flow or calibration due
date.
Home Screen Sample Map View
Home Screen Sample Map View