How to troubleshoot lack of handpiece power in Linvatec Hall 4 Medical Equipment?
- JJoseph ChungAug 1, 2025
If the Linvatec Medical Equipment handpiece lacks power, make sure the tank valve is completely open and that nitrogen is being used. Also, ensure the mode selector ring is in the proper operating position. If the issue persists, verify all hose connections are completely seated and remove any hose restrictions. Check if the tank pressure is below 500 psi and replace the tank if necessary. Finally, set the pressure to the recommended operating pressure. To determine the cause, run the handpiece on another regulator; if it works, replace the malfunctioning regulator.