What to do if ventilation pressure is implausibly high with Lowenstein Medical Equipment?
S
Sierra ShawAug 7, 2025
If the ventilation pressure is implausibly high, it may be due to accessories increasing the flow resistance. Try exchanging the breathing circuit or filter. Also, remove or prevent condensate in the breathing circuit.
C
Carolyn OconnorAug 12, 2025
How to fix a Lowenstein Medical Equipment when a USB stick is not detected?
K
Kenneth MillsAug 12, 2025
If the USB stick is not being detected, the issue might be incorrect formatting. The elisa 300 / elisa 500 devices only support a FAT32-formatted USB stick. Ensure the USB stick is formatted with a standard operating system like Windows.
A
austin37Aug 18, 2025
What to do if connected accessories don't work with Lowenstein Medical Equipment?
F
ftaylorAug 18, 2025
If a connected accessory isn't working, the function may not be configured for the user or it may have been configured for another interface. You can review the current interface configuration on the “Status” tab by selecting the second view.
S
Shawn ChambersAug 25, 2025
What causes a different measured PEEP value from the set value in Lowenstein Medical Equipment?
K
Kristen DavisAug 25, 2025
A difference between measured and set PEEP values can be caused by contamination of the PEEP membrane with medication residue after a drug nebulization procedure. This can happen if the expiratory filter is missing or unsuitable, or if the replacement interval is too long. To resolve this, exchange the expiratory valve and install a suitable expiratory filter for drug nebulization, or choose a shorter replacement interval.
K
Katherine MurphyAug 30, 2025
How to display a required reading on Lowenstein Medical Equipment?
C
Colleen JohnsonAug 30, 2025
If a required reading, ventilation mode, function, parameter, or curve is not displayed, it is possible that the desired function or display has not been configured for the user. Please contact the responsible person in your ward or your service technician.